Every time you exclude an endpoint, rely on existing data, claim biological equivalence, or choose which reprocessing timepoints to evaluate, you are making a judgment call that needs to be documented. Weak justifications are among the most common causes of deficiency findings in biological evaluation documentation. The problem is rarely that the decision was wrong. It is that the documented rationale is too vague or too generic to be evaluated by a reviewer with no prior knowledge of your device.
This article covers the anatomy of a strong justification, the most common deficiency patterns, and where FDA and EU notified body expectations differ.
What a justification actually needs to do
A justification explains why a specific decision is scientifically supportable for this specific device. A justification that would be equally valid for any device, regardless of its materials or clinical use, is not specific enough. ISO 10993-1:2025 requires the BER to document rationale for decisions at each stage, aligning with ISO 14971's traceability requirements.1 The EU MDR requires manufacturers to demonstrate through their GSPR documentation that biological risks are minimized, with notified bodies expecting traceable reasoning for every significant decision.2 The FDA similarly emphasizes that deviations from the evaluation framework should be justified with clear conclusions and identified gaps.3
The four components of a strong justification
1. What decision was made
State the specific decision clearly. "Genotoxicity testing was not performed." "Chemical characterization data from the predicate device was used in lieu of new extractable testing." "Biocompatibility evaluation was performed at T-end only, without baseline testing." The reviewer needs to know exactly what you decided before they can evaluate whether the reasoning supports it.
2. Why the decision is appropriate for this device
This is where most justifications fail. The rationale must connect the decision to the specific characteristics of the device under review: its materials, its contact type and duration, its reprocessing method, its clinical use, and its patient population. A statement that "the material has a long history of safe use" is not a justification because it does not identify which material, in what configuration, under what conditions, or reference any evidence.
3. What evidence supports the decision
Every justification must be traceable to specific evidence. This could be a test report, a published study, chemical characterization data, a toxicological risk assessment, a biological equivalence evaluation, or a documented risk assessment. The reviewer should be able to locate and evaluate the referenced evidence. If the evidence is "internal knowledge" or "industry practice," it is not traceable and will not hold up.
4. Why the evidence is still relevant
If the evidence comes from a prior evaluation cycle, a different device variant, or a predicate device, the justification must address whether anything has changed that would affect the relevance of that evidence. Material changes, manufacturing process changes, sterilization method changes, or changes in the clinical use of the device can all invalidate previously adequate evidence.
The most common justification failures
Generic material safety claims
"Silicone is a well-established biocompatible material widely used in medical devices. Additional testing is not required."
"The device body is manufactured from medical-grade polydimethylsiloxane (PDMS), Dow Corning QP1-40. Chemical characterization (Report CC-2024-017) identified no extractable compounds above the AET. TRA (Report TRA-2024-022) confirmed margins of safety exceeding 100 for all identified constituents. Cytotoxicity, sensitization, and irritation testing were performed on the final finished device (Reports CT-2024-008, SN-2024-011, IR-2024-014). Results for all endpoints were within acceptance criteria."
The weak version makes a claim about a general material class. The strong version identifies the specific grade, references specific test reports, and connects each conclusion to documented evidence. A reviewer can verify every claim in the strong version. They cannot verify anything in the weak version.
Circular reasoning
"Systemic toxicity testing is not required because the device does not present a systemic toxicity risk."
"Systemic toxicity was evaluated through chemical characterization (ISO 10993-18) and toxicological risk assessment (ISO 10993-17). Extractable testing (Report CC-2024-017) identified four compounds above the AET. TRA (Report TRA-2024-022) calculated margins of safety for all four, with the lowest MOS being 250 based on published NOAEL data. On this basis, the systemic toxicity risk is adequately addressed without additional in vivo testing."
The weak version uses the conclusion as its own justification. The strong version shows the work: it identifies the evaluation method, references the data, and explains how the conclusion was reached.
Undocumented precedent
"This approach has been accepted in previous submissions."
"This evaluation approach is consistent with the methodology applied to [Device X] (BER-2022-005), which was reviewed and accepted by [notified body/FDA] in [date]. The materials, contact type, and clinical use of the current device are identical to [Device X], and no changes have been made to the manufacturing process or sterilization method since the prior evaluation."
Prior acceptance is not inherently invalid as supporting evidence, but it must be documented with specifics. Which submission? Which device? What was the same, and what has changed? Reviewers change, expectations evolve, and standards are revised. An approach accepted in 2019 may not be accepted in 2026, especially following the 2025 revision of ISO 10993-1.
Missing rationale for excluded effects
One of the most frequent deficiency patterns is a BEP that lists which biological effects were evaluated and which were not, without documenting why the excluded effects were excluded. The 2025 standard's restructured evaluation tables make this more visible: if an effect is indicated for your device category, its absence from your evaluation requires a documented justification. For detailed guidance on this specific topic, see how to justify excluding biocompatibility endpoints in your BEP.
Where FDA and EU MDR expectations actually differ
Both sides want documented, evidence-based justifications. But the mechanics are different in ways that affect how you write.
The FDA gives you a checklist to work from. The FDA-modified biocompatibility matrix tells you which endpoints to evaluate based on your device's contact type and duration. If your evaluation follows the matrix, the justification burden is straightforward. If you deviate, you justify the deviation against a published agency position. The FDA has also published specific methodological expectations for chemical characterization, AET calculations, and test article preparation, so reviewers are checking your work against concrete guidance documents.3
EU notified bodies do not have an equivalent matrix. They assess your BEP and BER as part of the full technical documentation review and judge whether your evaluation is adequate based on the ISO standard, MDCG guidance, and their own interpretation of current "state of the art." This gives you more flexibility in how you structure your evaluation, but less certainty about what will be accepted. Notified body reviewers also tend to scrutinize cross-document traceability more heavily: a justification that is scientifically adequate but contradicts a statement elsewhere in your technical file will generate a finding.2
The EU requires CMR substance justifications the FDA does not. Under MDR Annex I, Section 10.4.1, devices containing carcinogenic, mutagenic, or reproductive toxic substances above 0.1% w/w require a documented justification that no suitable alternative exists and that the benefit-risk ratio supports their use. This is an entire category of justification that does not have a direct equivalent in the FDA framework.5
Equivalence is harder in the EU. The FDA's 510(k) substantial equivalence typically focuses on demonstrating that your device is similar enough to a predicate. EU notified bodies apply a three-phase assessment covering technical, biological, and clinical equivalence (MDCG 2020-5), and specifically scrutinize whether manufacturing differences could affect the release characteristics of substances from the device.4
For a detailed breakdown of these differences with specific examples and deficiency finding patterns from each side, see FDA vs. EU MDR expectations for biological evaluation: where they actually differ.
Practical implication: If you submit to both markets, write to the EU standard first (generally more demanding for documentation completeness and traceability), then ensure FDA-specific expectations (the modified matrix, agency guidance on chemical analysis) are also addressed.
Justifications across different decision types
Endpoint exclusions: Every excluded biological effect requires device-specific rationale. See the detailed endpoint exclusion guide.
Use of existing data: Confirm the test was performed on a representative article, conditions remain relevant, and no device changes have occurred since testing. If the test used a prior standard revision, acknowledge this and explain why the data remains adequate.
Biological equivalence: Identify the comparator, document the comparison across all relevant dimensions, and address any differences. Under the 2025 standard, focus on whether the biological risk profile is sufficiently comparable, not just whether inputs match. Under the EU MDR, additionally address MDCG 2020-5's three-phase framework.4
Test method deviations: Document what was different, why the deviation was necessary, and why results remain valid.
Reprocessing timepoints: Document why the selected timepoints adequately characterize the device across its reuse life. See the detailed reprocessing timepoint guide.
A self-check before submission
For every justification in your BEP and BER, confirm that the decision is stated clearly, the rationale is device-specific, the supporting evidence is referenced and traceable, and continued relevance is addressed for any prior data. Then check the one thing most teams miss: cross-document consistency. Does the rationale contradict anything stated in the BEP, BER, chemical characterization, TRA, or risk management file?
Cross-document consistency is one of the hardest things to maintain and one of the most common sources of deficiency findings. BioEvalPro scores every justification in your package for specificity, traceability, and internal consistency, and flags the ones that would not survive the scrutiny described in this article. request a gap analysis for early access, or reach out to discuss your documentation needs.
The principle: Reviewers do not penalize judgment calls. They penalize undocumented judgment calls. If a decision exists in your biological evaluation, a justification should exist alongside it. The quality of that justification determines whether the reviewer moves on or writes a deficiency.
1 NAMSA, "ISO 10993-1:2025 Updates: Top 10 Biological Evaluation Essentials in the Revision," October 2025. namsa.com
2 Regulation (EU) 2017/745 (Medical Device Regulation), Annex I (GSPR) and Annex VII (requirements for notified bodies, including competence verification for biocompatibility). See also Johner Institute, "Biocompatibility, Biological Safety, ISO 10993." johner-institute.com
3 FDA Guidance, "Use of International Standard ISO 10993-1: Biological evaluation of medical devices, Part 1: Evaluation and testing within a risk management process," updated 2023. fda.gov
4 MDCG 2020-5, "Guidance on Clinical Evaluation: Equivalence." See also MED Institute, "Challenges of Demonstrating Biological Equivalence per ISO 10993-18 and EU MDR." medinstitute.com
5 Regulation (EU) 2017/745, Annex I, Section 10.4.1 (CMR substance requirements). See also Sobel Consult, "Biological Evaluation Europe." sobelconsult.com
You now know what a strong justification looks like. How many of yours would pass?
BioEvalPro evaluates every justification in your package for specificity, traceability, and consistency. Generic claims, circular reasoning, and rationale that contradicts other sections of your documentation are flagged with specific guidance on what to fix.